TECHNICAL PARAMETER OF ANESTHESIA MACHINE PART

Console: high strength engineering plastic rack,light, beautiful and corrosion resistance
Scope: adult, child
Gas sourse: O2,N2O : 0.3-0.5 MPa
Flow meter: O2:0.1-10 L/min N2O:0.1-10L/min
O2,N2O linkage and N2O truncation device.When use N2O,O2 concentration>25%
When O2 pressure <0.2Mpa,the flow of N2O would be cut off.
Rapid oxygen flow: 35-75L/min
Low oxygen pressure alarm: it will be alarmed when the oxygen pressure <0.2 MPa
Vaporizer: it has automatic compensation function of pressure, temperature and flow automatic compensation. The regulation range of evaporator concentration is 0-5 vol%. Choose one among enflurane for a matching according to the demands of customer. Respiratory circuit:work mode all close,semi-close,semi-open.
APL: 1-4KPa
Bellows breathing: Adults and children bellows
Monitoring of humidity volume:0-1500ml


TECHNICAL PARAMETER OF VENTILATOR PART

Display mode: 5.1 inch LCD
Ventilation mode:IPPV,
SIPPV
Artificial monitoring manual(monitor the tidal volume,ventilation
Respiratory rate.ratio,peak pressure when use hand
PLV
PEEP
Ventilation parameters: Tidal volume 50-1500 ml
Frequency 6-60 times/min
Breath/expiratory ratio 2:1-1:4
Trigger sensitivity -0.2- -1.0 KPa
PEEP 0.2- 1.0 KPa
Pressure control 1 – 6.0 KPa
Ventilation monitoring parameters:
tidal volume,ventilation,IPPV ratio,total respiratory rate.Breath/expiratory ratio,peak pressure Pressure control,airway pressure wave.Breath pressure trigger.PEEP minimum pressure respiratory rate,breath/expiratory ratio,airway pressure Monitoring oxygen concentration: 21%-100%(choose and match)
Security alert system
Oxygen concentration alarm
Upper limit setting range 50%-100% Low limit setting range 15%-50%
Airway pressure alarm
Upper limit setting range 1.5-6.0KPa Low limit setting range 0.2-3.0KPa
Alarm minute ventilation
Upper limit setting range 2-30L/min Low limit setting range 1-12L/min
Asphyxia alarm the machine will give sound and light alarm if there is no tidal volume input in 15s
Power failure alarm
Power AC 220V±10% 50Hz±1Hz
UPS Battery(choose and match)


STANDARD CONFIGURATION

1.High-precision evaporator, either Halothane, Enflurane, Isoflurane or Sevoflurane configuration
2.O2 and N2O double-tubes flowmeter
3.Standard loop system
4.Modulated anaesthetic machine structure
5.Children or adult bellows
6.5.7″ LCD ventilator (IPPV, SIPPV, SIMV,PEEP, SIGH, MANUAL, ventilation volume, tidal volume, respiratory rate, peak pressure of airway)
7.Real-time display of pressure/time and flow-speed/time waves, incorporation of parameter setting and monitoring